MBP-TM series Bone Plate of Orthopedics Testing Machine
1 Important function
1) it is used in tensile, compression, bending test with synosteosis and bone fracture plate and other mechanic test and research.
2) medical instrument, synosteosis, surgical implant, straight bone fracture plate, angled bone fracture plate, grip type of bone fracture plate, pipe shape of bone fracture plate, L shape of bone fracture plate, hook bone plate, double bend bone plate, Y type of bone plate, golf stick shape of bone plate, T shape of bone plate, rebuild bone plate, cambered rebuild bone plate, wave form of steel plate, point shape of touching bone plate, touching with power pressuring bone plate.
3) device for bone fracture plate in marrow, bar, wire and board type of bone fracture plate, bone fracture plate fixed in thighbone, bone spicule and wire, metal bone spicule, forgeable metal wire, wrought wire for surgery department, metal bar for orthopedic, bone fracture plate for locking and compressing, steel plate for osteotomy, limited bone fracture plate for thighbone, near -end bone fracture plate for thightbone, bone fracture plate for anklebone, steel plate for upper limb, steel plate for lower limb, bone fracture plate of neck of femur, bone fracture plate for outside, limited touching bone fracture plate, supporting bone plate for condyles of femur, 3 leaf bone plate, suture needle with thread, steel wire for medical suturing.
4) Bending strength and rigidity, equivalent bending stiffness, loading - deflecting curve, tensile strength, yield strength, upper yield strength, lower yield strength, proof strength non-proportional extension, percentage elongation after fracture, percentage reduction of area, stress, tensile force, tensile to break, elasticity test, toughness test.
2 Standard
1) ISO 9585: Implants for surgery - Determination of bending strength and stiffness of bone plate
2) ISO 5832-12 - Implants for surgery-Metallic materials-Part12: Wrought cobalt-chromium-molybdenum alloy
3) ISO 13356-Implants for surgery-Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP)
4) NF S94-063 Implants for surgery-Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP).
5) ASTM F382-2014 Standard Specification and Test Method for Metallic Bone Plates
6) DIN ISO 5832-9 Implants for surgery -Metallic materials-Part9: Wrought high nitrogen stainless steel
7) EN ISO 5832-2-2012 Implants for surgery -Metallic materials. Unalloyed titanium
8) DIN ISO 5832-3- Implants for surgery -Metallic materials-Part3: Wrought titanium6-aluminium4-vanadium alloy
9) DIN ISO 5832-4- Implants for surgery-Metallic materials-Part4: Cobalt-chromium-molybdenum casting alloy
10) DIN ISO 5832-1-Implants for surgery - Metallic materials - Part 1: Wrought stainless steel
11) DIN ISO 5832-5-Implants for surgery - Metallic materials - Part 5: Wrought cobalt-chromium-tungsten-nickel alloy
12) DIN ISO 5832-6-Implants for surgery - Metallic materials - Part 6: Wrought cobalt-nickel-chromium-molybdenum alloy
13) DIN ISO 5832 – 7 - Implants for surgery - Metallic materials - Part 7: Forgeable and cold - formed cobalt – chromium – nickel – molybdenum - iron alloy.
14) DIN ISO 5832-8- Implants for surgery - Metallic materials - Part 8: Wrought cobalt-nickel-chromium-molybdenum-tungsten-iron alloy
15) DIN ISO 5832-9- Implants for surgery - Metallic materials - Part 9: Wrought high nitrogen stainless steel
16) DIN ISO 5832-11- Implants for surgery - Metallic materials - Part 11: Wrought titanium 6-aluminium 7-niobium alloy
17) ISO 6892-1-Metallic materials-Tensile testing-Part1:Method of test at room temperature
18) ASTM E8M-04.Standard Test Methods for Tension Testing of Metallic Materials
19)BS EN10002-1Tensile testing of metallic materials. Method of test at ambient temperature
20)ISO 7438:Metallic materials-Bend test
3 Important specification
No. |
Item |
Specification |
1 |
Load capacity |
50kN |
100kN |
200/300kN |
2 |
Grade |
Grade of 0.5 or 1 |
3 |
Range of measurement |
0.4% --100%FS(grade of 0.5) |
4 |
Relative error of indicated value |
Between ±0.5%(grade of 0.5) /±1.0%(grade of 1) |
5 |
Resolution of force |
1/±300000F.S(without change of whole range) |
6 |
Range of deformation |
0.2% -- 100%FS |
7 |
Range of force speed |
0.005% --5%FS/S |
8 |
Precision of force speed |
When speed is less than 0.05FS, between ±1% of setting value,
When speed is more than 0.05%FS, between ±0.5% of setting value. |
9 |
Range of deformation speed |
0.005 – 5%FS/S |
10 |
Precision of deformation speed |
speed is less than 0.55%FS/S, between ±1% of set value.
speed is more than 0.05%FS/S, between ±0.5% of set value. |
11 |
Range of displacement speed |
0.001 – 600mm/min (customized high speed system:0.001 – 2000mm/min) |
12 |
Precision of displacement speed |
between ±0.2% of setting value |
13 |
Range of constant force, strain, and displacement |
0.5% -- 100%FS |
14 |
Precision of constant force, strain, displacement |
set value is more than 10%FS, between ±0.5% of setting value.
set value is less than 10%FS, between ±1% of setting value. |
15 |
Test span |
400mm |
500mm |
610mm |
16 |
Crosshead stroke |
1000mm |
1000mm |
1250mm |
17 |
Voltage |
220±10%, 50Hz |
380±10%, 50Hz |
18 |
Power |
0.75KW |
2.0KW |
5.0KW |
19 |
Size |
940*520*1850mm |
1050*560*2150mm |
1050*750*2600mm |
20 |
Weight |
500kg |
800kg |
180kg |
|